Product Info
Pilot C is a syringe pump allowing IV infusion at low flow rate with accurate pressure monitoring. Pressure monitoring, performed via an electronic sensor, is permanently displayed and achieves a very short delay for occlusion detection.
Features:
- Flow rates: 0.1 to 800 ml/h
- Syringe selection 12 types of Luer lock 50/60 ml and 20 ml, 10 ml with accessory
- Flow rate accuracy ± 1% on drive mechanism/± 2% on syringes
- Name of the drug 15 drug names proposed, easy to customize
- Bolus rate programmable from 50 up to 800 ml/h, with infused volume display, protected access
- Two occlusion alarm programming modes: variable mode: numerically selected occlusion release (from 100 to 900 mmHg). Pre-selection mode: 3 preset pressure occlusion alarm levels (300/500/900 mmHg)
- Line disconnection alarm drop in pressure in the infusion line
- Anti-bolus system: bolus volume reduction at occlusion release
- Volume limit and total volume infused
- RS232 serial link for standard computer interface. TTL link for Master extension
- Battery life: 7 hours minimum, at 5 ml/h for a 50/60 syringe, displayed in hours/minutes
- Mains supply 12–15 V dc 15 W
- Dimensions/Weight W=120, H=330, D=155 mm/2.2 kg