A drug library including infusion protocols in mass units secures and eases IV infusion in the operating theatres.
Each drug protocol presets induction and maintenance.
Features:
- Choice of programming units, µg/ml or mg/ml, µg or mg per kg per hour or minute, ml or ml/h
- Drug library up to 29 pre-programmable agents. Protocols include default settings and min/max values for drug, drug concentration, induction dose and rate, maintenance rate, bolus parameters
- Master applications Master range for dedicated applications and evolution: Patient Controlled Analgesia with Master PCA. Target Controlled Infusion with Master TCI
- Syringe selection 25 syringes of 20 and 50/60 ml; 10 ml with accessory
- Accuracy ± 1% on drive mechanism/± 2% on syringes
- Flow rate selection range 0.1 to 1500 ml/h
- Pressure limit occlusion pressure alarm: from 100 up to 1100 mmHg. Automatic increase of occlusion threshold alarm at bolus injection
- Pressure drop detection: adjustable from 100 up to 1500 mmHg
- Anti-bolus system significantly reduces bolus after occlusion release
- History log last 3 patients: patient weight, used drug, concentration, infused dose, infused volume, infusion
- Computer connections RS232 serial link for standard computer interface/TTL link for Master extension
- Battery life: 7 hours minimum at 5 ml/h for a 50/60 syringe, displayed in hours/minutes
- Dimensions/Weight W=120, H=330, D=155 mm/2.2 kg